Injection

Positioning

The first step of the injection process is to select the positions where the injections should take place. Refer to Position Selection to see how to select positions for injection.

Injection

Injection Screenshot

Procedure

This step approaches towards the surface until the configured setpoint is reached and then keeps it for the configured amount of time by adjusting the z position. After this, the probe is retracted by a fixed distance.

The configured pressure is applied while the probe is in contact. Note that delays of one second are applied before and after the pressure pulse to achieve better results. This implies that the time during which the probe stays in contact is always two seconds longer than configured in the program.

Idle pressure is applied during approach and retraction.

Injection Screenshot

Review

Refer to Position Selection to see how to save the processed positions as a point group.