Stage XY

Preparation

General requirements

General installation conditions: for a secure installation of FluidFM Bot positioning system on the superior unit (e.g. optical microscope) it must be ensured:

Installation of the FluidFM XY

Unpacking of the FluidFM Bot positioning system

Open transport box and remove packaging.

Screenshot
FluidFM Bot XYZ positioning system in the transport crate

Lift the FluidFM XYZ positioning system from the crate

The positioning system weighs more than 25 kg.

Before lifting the system manually, the Fluid FM Bot Z-axis (8.5 kg) must be first disassembled from the FluidFM Bot XY-axis (16.5 kg) and be taken out separately.

Warning!

Screenshot

Never lift the FluidFM Bot XY-axis and FluidFM Bot Z-axis by sheet metal covers! Lift the axis by gripping on structural frame only. Never lift the FluidFM Bot XY-axis- and FluidFM Bot Z-axis with the cables! Avoid deformity and mishandling of the cable to prevent cable damage! Damaged cables may cause malfunctions!

Screenshot
FluidFM XY-stage
Screenshot
FluidFM Bot XY-stage transport safety (fitting screw)

Remove transport safety fitting screw, as showed in the video:

Installation of FluidFM Bot XY

Mounting position of the FluidFM Bot XY-stage in the superior unit (e.g. optical microscope) is horizontal.

Requirements for installation: interface for installation on the superior unit:

Stage XY is mounted with the screws onto the microscope. Mounting surface must maintain a flatness minimum 15 µm.

Depending on the microscope’s type must be used spacers or distance blocks. The assembly of FluidFM Bot XY-axis onto the Zeiss Observer and Olympus microscope is described below. The Nikon and Leica are not described in this manual.

It must be ensured that superior unit is disconnected from the main line during the XY-Stage installation. To ensure proper grounding of the FluidFM Bot XY-stage together with superior unit.

The support bolts stabilize the XYZ-System on the microscope.

Assembling and adjusting of the support bolts:

Caution!

Screenshot

Too high bolts tension may negatively impact geometric accuracy of the XY-Stage.

Zeiss Observer microscope

Screenshot
Interface for Zeiss Observer. Fastening with 3xM4x16 screws

(Vorderseite XY-Stage is Vorefront XY-Stage):

Olympus IX83 microscop

Objectives Collision Danger

When operating the system with the Olympus IX83 microscope, always set the objective’s Escape Length to 9mm and the maximum z-focus travel range to 9.5mm to avoid collision of the objectives with the stage

Screenshot
View from below: distance blocks for Olympus microscope
Screenshot
View from above: tightening with 3xM4X12 screws