Cell Isolation
Procedure
Pick
This step approaches towards the surface and applies the configured vacuum once the probe touches the selected cell. After the configured duration expired the system retracts from the surface again.
After this step the cell should have been picked up.
Place
The deposition step performs a pre-release of the cell and then approaches to the surface according to the configured parameters. This is done at the current xy position. The probe hovers slightly above the surface for the configured duration and then retracts by the configured distance. An additional lateral motion is applied to make it simpler to visually confirm that the operation succeeded.
Parameters
Parameter | Description |
---|---|
Setpoint |
The probe deflection value at which the approach will be stopped |
Duration |
How long the probe remains on the surface before pick-up |
Pressure |
The pick-up pressure to apply after initial contact |
Deposition Contact Force |
The probe deflection value at which the approach will be stopped |
Release Pulse Time |
How long Release Pulse Pressure is applied |
Release Pulse Pressure |
Pressure to apply for pre-releasing the cell before approach |
Surface Contact Time |
Duration during which the probe hovers above the surface, before Retract Distance is applied |
Retract Distance |
Final retraction distance to apply at the end of placement |